Abstract

Effective dose (E) in medical procedures is of practical value for comparing doses from different types of procedures which is not possible with the different measurable dose quantities. In this survey we estimated and compared the values of E in 12 medical imaging procedures. The mean E for conventional X-ray in mSv were ranged between 0.01 for skull lateral to 0.5 for abdominal while the mean E for cardiac interventional radiology in mSv was up to 16 for percutaneous transluminal coronary angioplasty (PTCA). For dental radiology the mean E in μSv was 1.2 for intraoral and 10 for panoramic. In mammography the mean E was 0.5 mSv. Compared with the literature, chest postero–anterior (PA), lumbar spine AP, PTCA and mammography procedures had shown larger E values. The obtained results can help in justification of techniques.