Abstract

A type-A uncertainty analysis of the magnitude of the local gravitational field can be used to validate type-B uncertainty estimates of calibrations of triaxial accelerometers by rotation in the gravitational field. We demonstrate the application of this method to validate a type-B uncertainty estimate at the sub 0.1% level of the errors introduced by a high-accuracy, two-axis, accelerometer-characterization system.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call