Abstract

The performance of polyester film was examined as a disposable cover of the sample-carrier (silicon wafer). The cover offers convenience and economy for the total reflection X-ray fluorescence (TXRF) analysis of solution samples. Moreover, the measurement of Al Kα (1.487 keV) and P Kα (2.013 keV) peaks is possible because of the reduction of Si Kα X-rays (1.740 keV) from the sample-carrier. Not only the trace elements (Cr, Mn, Co, Ni and Cu) but also the low sensitivity elements (Al and P) in JSS iron and steel standard samples were successfully analyzed.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.