Abstract

This prospective, randomized protocol evaluated femoral head penetration after total hip arthroplasty in a young population. Forty-five patients randomly received either a cross-linked or conventional ultrahigh-molecular-weight polyethylene (UHMWPE) liner in a noncemented hemispheric cup (Trilogy, Zimmer, Warsaw, Ind) with a 28-mm femoral head. Radiostereometric analysis film pairs, Harris hip, UCLA, SF-12, and Western Ontario and McMaster Universities scores were obtained through 2 years. Median femoral head penetration was less among cross-linked compared to conventional liners as follows: 0.06 mm (0.04-0.08 mm) vs 0.08 mm (0.02-0.19 mm) at 6 months, 0.07 mm (−0.14 to 0.16 mm) vs 0.11 mm (0.01-0.27 mm) at 1 year, and 0.065 mm (−0.04 to 0.193 mm) vs 0.169 mm (0.09-0.22 mm) at 2 years. Clinical outcomes were similar between the groups. Highly cross-linked UHMWPE demonstrated 55% less femoral head penetration compared to conventional polyethylene at 2 years. Despite improvements in the manufacturing process and sterilization of conventional UHMWPE, the femoral head penetration rate is unchanged from historical standards.