Abstract
We compare calibrations for use on three-sampler vector network analyzers (VNAs), which do not allow the direct application of some advanced error-correction schemes such as TRL (thru-reflect-line). Here we compare various alternatives, including an approximate version of TRL that has been introduced commercially and two-tier multiline TRL using external software. We consider both coaxial and coplanar open-short-load -thru (OSLT) calibrations for the first tier, showing that the latter can lead to inaccuracies. Finally, we investigate the stability of the load reflection terms to show that the first tier calibration need not be frequently repeated.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.