Abstract

A two-step sharpening process for a Si pillar in a quartz tuning fork sensor was developed for use in frequency modulation atomic force microscopy (FM-AFM). By combining electrochemical etching in HF solution and anisotropic etching in KOH solution, the tip radius was decreased to 120 nm. The Si-probe-attached sensor showed a higher resonance frequency than a tungsten-probe attached sensor, which would lead to a higher force sensitivity. We demonstrated FM-AFM imaging on a cleaved mica surface in aqueous solution by using the fabricated sensor, and atomic resolution was successfully achieved.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.