Abstract

Electrocardiography (ECG) signal is currently recorded using minimum three electrodes, two electrodes for differential measurements and the third one to increase the Common Mode Rejection Ratio (CMRR). This paper presents a method to increase the CMRR of the analog front-end of ECG circuits. It is shown that the CMRR of ECG systems can be improved significantly without the need to the third electrode using a feed-forward method. It is also shown that the implemented ECG circuit can outperform the conventional right-leg driven systems when the signal paths are matched. The presented scheme can also be used with the right-leg driven circuit to further improve the CMRR.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.