Abstract

Two-dimensional Fourier transform ion cyclotron resonance (2D FTICR) mass spectrometry is a developing form of data-independent acquisition that allows for the simultaneous fragmentation and correlation of fragment ions to their precursors across a range of m/z values. The modern usage of 2D FTICR is performed using electrospray ionization (ESI) as the dried droplet preparation for matrix-assisted laser desorption ionization (MALDI) does not produce a consistent packet of ions over a number of scans. This work uses pneumatic spray techniques from mass spectrometry imaging to create a homogeneous surface for use with MALDI as an ionization source for 2D FTICR. A mixture of peptides and matrix was deposited onto a glass slide using an HTX pneumatic sprayer. MALDI was then used to ionize the peptide mixture for use with a standard 2D FTICR pulse sequence. The generated 2D spectrum reveals comparable structural information to spectra collected in a 1D experiment. Artifacts observed in the collected 2D MALDI spectra do not significantly differ from those expected from 2D ESI spectra.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.