Abstract
Two-dimensional Fourier transform ion cyclotron resonance (2D FTICR) mass spectrometry is a developing form of data-independent acquisition that allows for the simultaneous fragmentation and correlation of fragment ions to their precursors across a range of m/z values. The modern usage of 2D FTICR is performed using electrospray ionization (ESI) as the dried droplet preparation for matrix-assisted laser desorption ionization (MALDI) does not produce a consistent packet of ions over a number of scans. This work uses pneumatic spray techniques from mass spectrometry imaging to create a homogeneous surface for use with MALDI as an ionization source for 2D FTICR. A mixture of peptides and matrix was deposited onto a glass slide using an HTX pneumatic sprayer. MALDI was then used to ionize the peptide mixture for use with a standard 2D FTICR pulse sequence. The generated 2D spectrum reveals comparable structural information to spectra collected in a 1D experiment. Artifacts observed in the collected 2D MALDI spectra do not significantly differ from those expected from 2D ESI spectra.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.