Abstract
We have developed a turnkey, fast response two-channel NOx monitor (NOx&#160;=NO + NO2) which provides&#160; simultaneous measurements of both NOx and NO2 and thus NO by subtraction.&#160; The NOx channel employs a carefully controlled concentration of photolytically-produced ozone (O3) to convert NO into NO2.&#160; The monitor can measure NO2 and NOx with an accuracy of better than 5% and precision of < 0.2 ppb (1s, 1s) with <1 second physical time constant in each measurement channel.&#160; The monitor is based on Aerodyne Research&#8217;s patented CAPS (Cavity Attenuated Phase Shift) technology previously employed to measure NO2concentrations, aerosol optical extinction and aerosol single scattering albedo.&#160;A CAPS-based NO2 monitor utilizes a light-emitting diode (LED) as a light source (450 nm for the NO2 channel and 405 nm for the total NOx channel) and a sample cell incorporating two high reflectivity mirrors; a vacuum photodiode is used to detect the light emitted from the cell. The square wave modulated light from the LED passes through the absorption cell and is detected as a distorted waveform which is characterized by a phase shift with respect to the initial modulation.&#160;&#160; The amount of that phase shift is a function of fixed instrument properties - cell length, mirror reflectivity, and modulation frequency&#8211; and of the presence of variable concentrations of nitrogen dioxide.&#160; The mixing ratio is calculated from the value of the cotangent of the phase shift, the speed of light, LED modulation frequency and the absorption coefficient of NO2.&#160; The use of 405 nm detection in the total NOx channel greatly reduces the possible interference caused by the presence of several ppm of O3 in the sample flow.We present data from an intercomparison of the NOx monitor with Aerodyne TILDAS monitors measuring both NO and NO2.&#160; The TILDAS monitors utilize infrared diode lasers to probe individual ro-vibrational absorption lines and are considered the &#8220;gold standard&#8221; for measuring concentrations of small molecules.&#160; The monitors were deployed on the Aerodyne Mobile Laboratory during the MOOSE (Michigan Ontario Ozone Source Experiment) campaign in late 2021.&#160; The figure presented below compares the measurement of both NO2 and the sum of NO+NO2 measured by the TILDAS monitors compared with that obtained by the CAPS NO&#160;monitor at 1 second resolution.&#160; Note the excellent agreement with respect to both magnitude and time resolution.&#160; Long term comparisons indicate that correlation coefficients at 1 second&#160; resolution approach 1.&#160;
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.