Abstract

A two-wavelength transmissometer employing a He-Cd laser (lambda(1) = 0.325 microm) and a He-Ne laser (lambda(2) 3.39 microm) has been developed for measuring the Sauter mean diameter of mineral ash droplets in high-temperature high-velocity coal-fired combustion flows. From transmission measurements at the two wavelengths, it is shown that mean diameters in the 0.3-3.5-,microm range may be inferred with a weak sensitivity to particle refractive index and size distribution shape. The volume concentration or loading of the aerosol may then be determined from the measured transmission at either wavelength. The instrument has been used to measure the mean size and loading of ash droplets in a pulverized coal-fire channel flow at temperatures to ~2900 K and velocities of up to 400 m-sec(-1) for combustion MHD power generation applications.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.