Abstract
Introduction: T-wave oversensing (TWOS) is generally seen in patients with hypertrophic cardiomyopathy (HCM) and is a rare cause of inappropriate implantable cardioverter defibrillator (ICD) shocks. TWOS rarely causes pacemaker dysfunction. Case Presentation: In this paper, we present two patients with hypertrophic cardiomyopathy (HCM). One patient had several in- appropriate ICD shocks, and the other experienced pacemaker dysfunction due to TWOS. ICD interrogation revealed that TWOS occurred only during high heart rates in the first patient. Attempts to fix TWOS, including a higher beta blocker dose, electrophysi- ology study, and ICD re-programming, were unsuccessful. Conclusions: We replaced the previously implanted ICD generator (Medtronic Maximo II DR) with a new one that has a specific diagnostic algorithm to prevent TWOS (Medtronic Protector). After replacement, the patient did not have any inappropriate shocks due to TWOS. The second patient had bradycardic pacemaker rhythm due to TWOS. Although we reposed, the ventricular lead did not stabilize with an adequate threshold, and thus we decided to replace the lead with a new one. The pacemaker dysfunction disappeared after the lead replacement. Patients with HCM must be observed carefully for these undesired conditions.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.