Abstract

A microtrap is a small diameter sorbent trap packed with adsorbent. In previous studies it has been used as a concentration cum injection device for on-line GC monitoring of trace level organics. The microtrap is prone to breakthrough as it contains a small quantity of adsorbent. A larger diameter trap with more adsorbent reduces breakthrough, but generates broad injection bands that reduce chromatographic resolution. In this paper, we present a two-stage microtrap system. The first, a large diameter retention trap is packed with more sorbent material to increase the breakthrough volume or time. The retention trap is desorbed and the analytes are refocused onto the smaller diameter microtrap referred to as the injection trap. A few seconds delay is provided and then the injection microtrap is desorbed to generate a sharp band injection for GC separation.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.