Abstract

Two RP-HPLC with fluorescence detection alternative methods have been proposed to quantify and identify bisphenol A diglycidyl ether (BADGE), a monomer of epoxy resins used as a coating for food packaging materials, and its hydrolysis products (HPs) in three aqueous food liquid simulants: 15% (v/v) ethanol, 3% (w/v) acetic acid, and distilled water. Both isocratic and gradient chromatographic methods were developed to determine this monomer at the specific migration limit (SML) level (restriction proposed by European Union legislation) and also the two HPs in aqueous EEC food simulants. All calibration lines had correlation coefficients greater than 0.997 and detection limits of less than 2 Ig of BADGE/L. Standard deviations (in percent) obtained in the precision calculation were less than 3% for BADGE. Furthermore, both methods were applied to analyze these compounds in three types of epoxy can coatings that contact food.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.