Abstract

An adaptive pulse shaper controlled by multiphoton intrapulse interference phase scanning (MIIPS) was used, together with a prism-pair, to measure and cancel high-order phase distortions introduced by a high-numerical-aperture objective and other dispersive elements of a two-photon laser-scanning microscope. The delivery of broad-bandwidth (approximately 100 nm), sub-12-fs pulses was confirmed by interferometric autocorrelation measurements at the focal plane. A comparison of two-photon imaging with transform-limited and second-order-dispersion compensated laser pulses of the same energy showed a 6-to-11-fold improvement in the two-photon excitation fluorescence signal when applied to cells and tissue, and up to a 19-fold improvement in the second harmonic generation signal from a rat tendon specimen.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.