Abstract
A ferromagnetic T-fastener is often used in children during gastrostomy procedures. We describe the experience with the use of a new gastropexy device (two parallel needles with a snare) that uses a non-ferromagnetic anchor during primary percutaneous retrograde radiologic gastrostomy insertion in four children with a mean age of 6years and 6months (range: 2years and 11months to 9years and 6months). The new gastropexy device was safe and effective in children. It is advantageous for children requiring magnetic resonance (MR) examinations soon after gastrostomy insertion, especially in view of MR-related implications that result from the use of a ferromagnetic T-fastener.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.