Abstract

Solid contact ion selective electrodes are extensively utilized owing to their marvelous performance over traditional liquid contact ones. The main drawback of those solid contact electrodes is aqueous layer formation which affects their constancy. Herein and to overcome this common drawback, a carbon paste electrode containing poly(3,4-ethylenedioxythiophene) was constructed and used for determination of probenecid at variant pH values. This modification decreased the potential drift down to 0.8 mV/h and improved its stability over 30 days. A Nernstian slope of − 57.8 mV/decade associated with a linear range of 1.0 × 10−6–1.0 × 10−2 mol/L was obtained. The modified carbon paste electrode successfully detected up to 8.0 × 10−7 mol/L probenecid. Results of this modified carbon paste electrode were also compared to unmodified one.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.