Abstract

We describe the design and characterization of a high resolution, high efficiency detector for two-dimensional imaging of neutral atoms. Incident atoms are surface-ionized by a tungsten-coated hot ribbon and the resulting ions are accelerated into a microchannel plate detector with a phosphor screen output. With this design we can detect individual alkali and other low ionization potential atoms and molecules with a spatial resolution of ∼20 μm. We find backgrounds on the order of 30 Hz over an active detection region of 20 mm2, a time response of ⩽1/30 s, and a detection efficiency of ∼50%. This detector can be used to image cold as well as thermal atomic or molecular beams.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.