Abstract

A two-dimensional (2-D) blood velocity estimator is presented combining speckle tracking (ST) and phase-shift estimation (PE) to measure lateral (vx) and axial (vz) velocities respectively. Estimator properties were assessed in a carotid bifurcation using ultrasound simulations based on computational fluid dynamics, allowing validation toward a ground truth. Simulation results were supported with in vivo data of a healthy carotid. ST and PE estimates were combined as: (1) vx from 2D-ST and vz from PE, (2) vx from 2D-ST and vz from PE with aliasing correction based on ST and (3) vz from PE and only lateral ST for vx. Regression analysis showed a 35% to 77% decrease in standard deviation for vz for PE compared with ST. Aliasing correction based on ST improved results but also introduced spurious artifacts. A marginal decrease in performance was observed when only tracking laterally. Further work will focus on in vivo trials in patients with carotid plaques. (E-mail: abigail.swillens@ugent.be)

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.