Abstract

A method for pesticide residue analysis applying the technique of two-dimensional gas chromatography with effluent splitting to three selective detectors in parallel is described. Screening and confirmation procedures are combined in one single instrument. The column combination used allows the separation and identification of more than three hundred pesticides on the basis of the resulting retention times and detector signals. Critical pairs of pesticides are distinguished by using the response ratio of the selective detectors characteristic for each substance. An example is given to demonstrate the merits of the described GC-system using the multimethod S19 of the DFG Pesticide Commission for clean-up.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.