Abstract

Percutaneous patent foramen ovale (PFO) closure is recommended for secondary prevention of paradoxical embolism through a PFO. In the United States, two Food and Drug Administration-approved PFO closure devices are currently available, and the choice depends on operator preference and PFO anatomy. Although these devices are easy to implant, there are several potential complications. As opposed to the Amplatzer PFO Occluder, there has been no published case of atrial erosion with Gore closure devices. This report describes two cases of pericardial tamponade due to perforation of the atrial wall induced by a wire frame fracture of the Gore Helex and Cardioform devices.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.