Abstract

Serum bone alkaline phosphatase (ALP; EC 3.1.3.1) was measured with a wheat germ agglutinin (WGA) precipitation assay and with a new IRMA in a group of healthy elderly women. Both assays were correlated with serum total ALP activity and with osteocalcin. The two bone ALP assays have comparable within- and between-run imprecisions (WGA assay within-run CVs 2.6-5.4% and between-run, 4.0-5.1%; IRMA within-run CV 5.0% and between-run, 3.2%). Comparison of the WGA precipitation assay (x) with the IRMA (y) demonstrated a correlation coefficient of 0.87 [Deming regression equation: y = (0.58 +/- 0.02)x - (4.62 +/- 0.45); n = 101; Sy/x = 1.26; P < 0.001). Correlation studies with osteocalcin and total ALP showed correlation coefficients (all P < 0.001) of 0.34 and 0.65, respectively, for the WGA precipitation assay and of 0.36 and 0.68, respectively, for the IRMA. We conclude that the two bone ALP assays have similar imprecision and that neither can be given preference over the other as a marker of bone turnover.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.