Abstract

The purpose of this study was to compare component alignment using two-dimensional (2D) and three-dimensional (3D) measurements following total knee arthroplasty (TKA) with a new robotic system. A total of 38 patients underwent TKA with the robotic system. Cutting errors were measured. Errors in 2D and 3D component alignments from the planned angle, validated cutting angle, and validated implantation angle were calculated. These errors were compared with 2D and 3D measurements. Cutting errors were below 0.6°, except for femoral sagittal angle (1.0°). Errors in component alignment from the planned and cutting angles were significantly higher in 3D measurements than that in 2D measurements, in the femoral coronal and sagittal planes and tibial coronal planes. Errors in component alignments from implantation showed no differences between the measurements. The accuracies of component alignment from the planned and cutting angles were better in 3D measurements.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.