Abstract

The simultaneous occurrence of multiple respiratory viruses, such as severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2 [CoV]) and influenza A virus (IAV), presents a challenge for accurate diagnosis and treatment. To address this, we developed the TwinDemic Detection (TDD) system—an integrated point-of-care diagnostic platform that can simultaneously detect CoV and IAV using a multi-respiratory virus identification approach. The TDD system contains a transparent poly(methyl methacrylate) microfluidic chip with a hydrogel-based gene detection sensor and a handheld fluorescence reader. This system uses a fuel-stimulated DNA probe to detect viral RNA and amplify the fluorescence signal without the need for enzymatic or thermal controls. The system’s performance was tested using 45 human nasopharyngeal samples; results revealed positive and negative predictive values of 93.3 % and 96.7 % for CoV, and 100 % and 96.7 % for IAV, respectively. This system enables the accurate and simultaneous detection of multiple respiratory viruses for public health surveillance and shows potential for use as a point-of-care molecular test to monitor a wider range of viruses in the future.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.