Twiddler's syndrome is a rare complication seen in patients with implanted pacemakers or defibrillators. The condition typically presents with device malfunction and occurs when the patient either consciously or unconsciously twists and rotates the implanted device in its pocket, resulting in torsion and dislodgement of the implanted lead. A case of twiddler's syndrome involving a transvenous defibrillation lead and an abdominally implanted defibrillator is described. This is the first report of this complication with this particular lead. The patient in this report was a middle-aged obese diabetic woman who presented 7 months after defibrillator implantation with device noncapture and intermittent nonsensing. Review of the literature reveals that the majority of patients with this complication are middle-aged obese women with a defibrillator pocket that exceeds the size of the defibrillator. Treatment measures are discussed both for the patient with this complication and for the patient at increased risk for its occurrence.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.