Abstract

A bolometer array diagnostic has been developed for the University of Washington Translation, Confinement, and Sustainment (TCS) field-reversed configuration (FRC) experiment in order to measure radially resolved total radiated power per unit length of the FRC. This will provide radiation energy loss information, useful in power balance and impurity studies. The 20-element photodiode bolometer will be mounted at the midplane of the TCS cylindrical vacuum chamber to view the rotating magnetic field (RMF) generated FRC plasma. Key features of this new bolometer array are (1) extensive electrical shielding against the RMF, (2) robust electrical isolation, (3) trans-impedance amplifiers using a microcoax interface at the array and a fiber optic interface to the screen room, and (4) a custom glass-on-metal socket for the 20-element photodiode chip to ensure high vacuum compatibility. The bolometer array can be retracted behind a gate valve using a stepper motor to protect it during vacuum chamber bakeout. The slit assembly housing is interchangeable to provide flexibility for the viewing sightlines.