Abstract

Compound-specific isotope analysis (CSIA) can provide unique insights into the cycling of elements including carbon and nitrogen. One approach for CSIA is the use of high-performance liquid chromatography (HPLC) to separate compounds of interest, followed by analysis of these compounds using an elemental analyser coupled to an isotope ratio mass spectrometer. A key component of this technique is the fraction collector, which automatically collects compounds as they are separated by HPLC. Here, we present a fraction collector that is a simple adaptation of a 3D printer, and, thus, can be easily adopted by any laboratory already equipped for HPLC. In addition to the much lower cost compared to commercial alternatives, this adaptation has the advantage for CSIA that the 3D printer is able to heat the collected fractions, which is not true for many commercial fraction collectors. Heating allows faster evaporation of the solvent, so that the dried compounds can be measured by EA–IRMS immediately. The procedure can be repeated consecutively so that diluted solutions can have the compounds concentrated for analysis. Any computer-controlled HPLC can be integrated to the fraction collector used here by means of AutoIt.