Abstract

Turbulent flow chromatography (TFC) is presented as a means to reduce ion suppression in simultaneous multianalyte mass spectrometric bioassays. In this study, the effects of enzymes present in the sample on the signal response of five analytes were simultaneously investigated over a protein content range from 0 to 38 microg/mL by means of direct flow injection MS. As model enzymes, trypsin, thrombin, and chymotrypsin were selected. Without employment of TFC, both signal suppression and signal enhancement, depending on the nature of the analyte and the amount of matrix in the sample, were observed. Generally, these matrix effects were found to be intolerably large. The deviation from the mean signal response as a measure of deterioration was found to be between 14 and 112%. The addition of an excess of methanol as means of sample clean-up was investigated and found not to be sufficient. By employing TFC for online sample preparation, it was possible to reduce the matrix effecTs to a minimum for all model systems investigated. In case of trypsin the distortion could be lowered from 41.9 to 2.6%. Thus, TFC is considered to be a highly valuable tool for improving the sensitivity and reliability in the monitoring of enzymatic conversions by means of MS.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.