Abstract

Both turbidimetric and photometric determinations of total tannins in samples of green and black tea, using a micro-flow-batch analyzer (μFBA) were studied. The miniaturized system was formed using photocurable urethane-acrylate resin and ultraviolet lithography technique. The turbidimetric method was based on the precipitation reaction of Cu (II) with tannins in acetate medium at a pH of 4.5. The photometric method was based on the complexation reaction of tannins with ferrous tartrate. The turbidimetric μFBA was able to test 200 samples per hour. The photometric μFBA allowed 300 analyses per hour, generating 136μL of residue per analysis. The paired t test, at a 95% confidence level, showed no statistically significant differences between results obtained by both methods and the reference method. The urethane-acrylate μFBA maintained satisfactory physical and chemical properties, and represents an improvement over conventional flow-batch analyzer.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.