Abstract

We describe a low field, bench-top MRI system for small samples, based on a permanent magnet. Signals are received at 830 kHz using a tuned SQUID magnetometer, cooled in a modified liquid helium cryostat. The SQUID input circuit has an intrinsic Q-factor of 28,000, so this is damped by flux-locking electronics and an additional feedback loop to give an effective Q of between 40 and 200 for imaging. The resonant frequency is adjusted by a control rod coupled to a trimmer capacitor mounted in the liquid helium volume. Images of a test object acquired using the cooled receiver exhibit up to two-fold SNR gains in regions close to the sensor, compared to an equivalent room temperature coil, with the noise level dominated by losses coupled from the magnet pole faces.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.