Abstract
The development of a tunable diode laser system for the puff-by-puff determination of hydrogen cyanide is described. Both spectral line scanning and fixed-frequency monitoring techniques are utilized for quantitative measurements on mechanically diluted cigarettes. It is shown that such mechanical dilution leads to significantly greater reductions in hydrogen cyanide delivery than expected from static dilution measurements.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.