Abstract

This work describes a methodology for Cd and Pb determination in sewage sludge slurry samples using thermospray flame furnace atomic absorption spectrometry (TS-FF-AAS). The TS-FF-AAS system was equipped with a Ni tube atomizer placed on an oxidizing air/acetylene flame. Two multivariate calibration models based on partial least squares (PLS) were proposed using total peak profiles (recorded during 57 s). Metals concentration in the sewage sludge samples were from 1.87 to 6.26 mg kg −1 for Cd and from 101 to 327 mg kg −1 for Pb. The limits of detection and quantification were, respectively, 0.2 and 0.7 μg kg −1 for Cd and 8 and 26 μg kg −1 for Pb. These values were three times lower than the limits found when these metals were calibrated using linear calibration with aqueous standard solutions.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.