Abstract

The “true detection limit” CL,true was measured by a method proposed by Boumans et al. for 10 prominent lines of La, Ce, Pr, Nd and Sm, the most abundant rare earth elements (REEs) in geological samples. It is demonstrated how spectral interferences can increase of CL,true. The spectra were excited in an argon arc plasma jet with evaporation powders of five complex natural samples. The data show that there is a significant increase of CL,true in real samples containing large concentrations of such elements as Ti, Zr, Cr, Nb, Ta and REEs emitting complex spectra.