Abstract
In this study, it has been shown that the true coincidence summing correction factor can be obtained for the first time using the PHITS Monte Carlo program. Determining this correction factor using different methods and tools in each laboratory to increase the possibility of achieving high-efficiency measurement conditions is still popular in gamma-ray spectrometry. By using 133Ba, 152Eu, 134Cs, and 60Co point sources, the true coincidence summing factor was investigated in both near and far counting geometries for 15 different energy values. GESPECOR software was used to validate the results obtained with PHITS. A remarkable agreement was obtained between PHITS and GESPECOR, with a discrepancy of less than 3%. With this study, a new tool has been proposed to obtain the true coincidence summing factor, which is one of the significant correction factors investigated/calculated in gamma-ray spectrometric studies.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.