Abstract

A personnel neutron dosimeter prototype based on chemical and electrochemical etched CR-39 detector, combined with LiF converter, has been calibrated using an ICRP-like phantom, under a heavy-water moderated Californium source neutron spectra; A conversion factor of 1.052±126 spots cm −2 mSv −1 was obtained. The sealing properties of the detector holder showed a ten-fold reduction in radon background when it was tested in a high radon atmosphere. A convenient mechanical shock resistance was achieved in LiF converters by sintering to 11 tons pressure LiF powder at 650°C, during one hour.