Abstract

We developed a high precision ionization chamber MALIC. It is an integral type ionization chamber which has a magnetically levitated electrode without being supported by insulation materials. Furthermore, the electric potential of that electrode can be read without physical contact, so that the contact noise does not occur. With MALIC, we attempted to measure the radiation dose from tritium vapor of a room air in an accelerator facility.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.