Abstract
The combustion method has been modified to increase the recovery of tritiated water after combustion of a tritium-labeled tissue sample. This was accomplished by cooling the bottom of the combustion flask in a dry ice-acetone bath while irradiating the top with an infrared lamp. The procedure resulted in at least 92% to 102% recovery of the tritiated water. The NCS solubilizer was found to be superior to hyamine for solubilizing(14)C labeled tissue samples. The samples yielded light yellow-colored solutions when incubated for 15 hr at 50-55C. The counting efficiency of this solution was 75% or higher.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.