Abstract

A common technique for the determination of the activity of tritiated water (HTO) is liquid scintillation counting (LSC). This implies sample taking, sample preparation and radioactive waste processing afterwards. When handling highly tritiated water special care has to be taken because of possible harmfulness of the sample. Furthermore, LSC devices are mostly large, heavy and expensive. The TRAMPEL experiment aims at measuring the activity of tritiated water in-line without sample taking. The device is intended to be easy to use and operate, quite inexpensive and compact. The measurement principle is based on electrons from β-decay which induce light in commercially available scintillation fibres. The light is detected by a small photomultiplier tube (PMT). A proof-of-principle was set up for static measurements using standard stainless steel parts. The complete device has a volume of less than 0.5 l.