Abstract

Successful long-term clinical results with the Charnley low-friction torque arthroplasty have identified proximal femoral strain shielding as a long-term problem to be addressed. The problem has arisen because of the load transfer by a distally supported stem that is no longer subject to fracture. To overcome this problem and improve proximal load transfer to the femur, a continuous triple-tapered stem—the C stem—was designed. This article describes the first 500 primary hip arthroplasties at a mean follow-up of 3 years, 5 months (range, 1-7 years). There have been no revisions for aseptic stem loosening, and no stem is considered to be at risk for loosening. In 20% of cases, there was subjective radiologic improvement of the bone–cement interface.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.