Abstract

In the TOMAS device, a triple Langmuir probe is used to measure the electron temperature and density. The accuracy of this measurement depends on correct determination of the effective collecting area of the probe, which depends on complex plasma transport processes. The probe can be calibrated by electron cyclotron resonance heating experiments using the cut-off density of the ordinary wave (O-wave). This threshold only depends on the frequency of the injected wave, and the occurrence of this phenomenon is clearly visible in the temperature evolution. The value of density is consequently known at this point and can be used to calibrate the density measurements of the triple Langmuir probe.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.