Abstract

Triboelectric charging is commonly detected during cardiopulmonary bypass in circuits using roller pumps and PVC tubing. Dissipation of this charging is needed to prevent a spontaneous discharge from occurring. We evaluated the ability of the Quadrox-D (Maquet) to effectively remove the electrostatic charge accumulation, with and without a heater/cooler connection (H/C). A Quadrox-D oxygenator was evaluated using a Stockert SIII pump head, Cincinnati subzero H/E, and a custom adult extracorporeal membrane oxygenation tubing pack with 1/2′ PVC raceway. The circuit was primed and evaluated for triboelectric accumulation with and without H/C use. The results showed a linear relationship between increasing pump flow and static charge build-up when an H/C was not applied. The calculated r2 value was .95. Incorporation of the H/C effectively eliminated charge accumulation. Increasing pump speed increases the amount of static charge created without the use of an H/C. Incorporation of an H/C effectively eliminates charge build-up in the Quadrox-D and is recommended while priming the circuit.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.