Abstract

Application of focused ultrasound stimulation (FUS) to the rat cervix during pregnancy has significant physiologic effects. One-millisecond-long pulses of 680-kHz ultrasound with a repetition frequency of 25 Hz, at ISPTA (spatial-peak, temporal-average intensity) of 1, 2 and 4W/cm2, were applied to the rat abdomen over the cervix. FUS produced a significant change in cervical elasticity known as softening, which is part of the ripening process, comparable to the degree seen just before delivery. Timed-pregnant Sprague-Dawley rats (n = 40) were used. During gestation, the FUS system was applied to the cervix for variable times up to 1 h. Daily measurements of cervix light-induced florescence were made to estimate changes in softening. In addition, cervical stretch estimates of softening were made of isolated cervices of control and FUS-treated rats to measure distensiblity. The ultrasound power with ISPPA (spatial-peak, pulse-average intensity) of 40 W/cm2 was considered tolerable; the U.S. Food and Drug Administration regulatory limit is 190 W/cm2 for both the body periphery and the fetus. This is the first report of alterations induced by ultrasound in the connective tissue of the cervix and suggests the therapeutic application of ultrasound for the facilitation of labor and delivery.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.