Abstract
Conventional planning of radiotherapy of tumours of the parotid, middle ear and other tumours in the head and neck often requires the treatment plane to be non-transverse. This produces major problems in delineating the tumour as well as verifying that vital structures such as the spinal cord are not included in the target volume. The use of computed tomography (CT) generally overcomes some of these problems and we have developed an algorithm to reconstruct transverse CT images into the appropriate oblique plane. Software has been written on the Picker Independent Treatment Planning System (ITPS) to allow planning on central axis and off-axis oblique planes. In addition we have used a beam's eye view facility to aid in the verification process.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.