Abstract

Blood bank regulations and bio medical waste rules of India advocate disinfection of contaminated blood units. Incineration is not recommended due to poly-vinyl chloride (PVC) content of blood bags. This study was designed to evaluate the efficacy of chemical disinfection of blood units deliberately contaminated with Staphylococcus aureus and E. coli with 1 and 6 % hypochlorite, 10% formalin and 33% formaldehyde and autoclaving of blood units contaminated with the above mentioned vegetative forms and B. stearothermophilus spores. Only 33 % formaldehyde could bring about 5 log reduction of bacteria but it is highly irritating and toxic. Autoclaving at 15 lbs pressure for 2 hours uniformly inactivated the vegetative forms and B. stearothermophilus spores. Thus, autoclaving of PVC blood bags is a safer and reliable method compared to chemical disinfection.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.