Abstract

A simple, portable and battery-powered trapping device (iTrap) consisting of a purification tube, a trapping unit and a miniature air pump was developed for the pre-concentration of volatile organic sulfur compounds (VOSCs). The tested VOSCs, including methanthiol (MT), ethanethiol (ET), dimethyl sulfide (DMS), diethyl sulfide (DES) and dimethyl disulfide (DMDS), were firstly purged from water samples and then in situ pre-concentrated with the iTrap prior to their analysis by thermal desorption gas chromatography coupling with a sulfur chemiluminescence detector (TD-GC-SCD). Twenty-six adsorbents were studied to find the most suitable adsorbent for the efficient pre-concentration of VOSCs. Under optimal conditions, limits of detection of 6, 8, 6, 2 and 3 ng L−1 were obtained for MT, ET, DMS, DES and DMDS, respectively. The precisions were better than 5.3% (relative standard deviations, RSDs). The iTrap was successfully applied for the analysis of VOSCs in Certified Reference Materials, several surface water, underground water and wastewater samples collected from Pengzhou city, Sichuan, China. Moreover, the VOSCs trapped in the iTrap were much more stable than those directly stored in water samples and the recoveries for all samples could be maintained at acceptable levels (>73%), even their preservation time as long as 8 h.