Abstract

A simple, portable and battery-powered trapping device (iTrap) consisting of a purification tube, a trapping unit and a miniature air pump was developed for the pre-concentration of volatile organic sulfur compounds (VOSCs). The tested VOSCs, including methanthiol (MT), ethanethiol (ET), dimethyl sulfide (DMS), diethyl sulfide (DES) and dimethyl disulfide (DMDS), were firstly purged from water samples and then in situ pre-concentrated with the iTrap prior to their analysis by thermal desorption gas chromatography coupling with a sulfur chemiluminescence detector (TD-GC-SCD). Twenty-six adsorbents were studied to find the most suitable adsorbent for the efficient pre-concentration of VOSCs. Under optimal conditions, limits of detection of 6, 8, 6, 2 and 3 ng L−1 were obtained for MT, ET, DMS, DES and DMDS, respectively. The precisions were better than 5.3% (relative standard deviations, RSDs). The iTrap was successfully applied for the analysis of VOSCs in Certified Reference Materials, several surface water, underground water and wastewater samples collected from Pengzhou city, Sichuan, China. Moreover, the VOSCs trapped in the iTrap were much more stable than those directly stored in water samples and the recoveries for all samples could be maintained at acceptable levels (>73%), even their preservation time as long as 8 h.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.