Abstract

PurposeThe main purpose of this work is the inter-comparison between different devices devoted to the transversal dose profile recostruction for daily QA tests in proton therapy. MethodsThe results obtained with the EBT3 radiochromic films, used as a reference, and other common quality control devices, have been compared with those obtained with a beam profiling system developed at the “Laboratori Nazionali del Sud” of Italian Institute for Nuclear Physics (INFN-LNS, Catania, Italy). It consists of a plastic scintillator screen (thickness 1 mm), mounted perpendicularly to the beam axis and coupled with a highly sensitive CCD detector in a light-tight box. Results and conclusionThe tests, carried out both at the INFN-LNS and Trento Proton Therapy Center facilities, show, in general, a good agreement between the different detectors. The beam profiling system, in particular, appears to be a promising quality control device for 2-D relative dosimetry, because of its linear response in a dose rate range useful for proton therapy treatments, its high spatial resolution and its short acquisition and processing time.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.