Abstract

BackgroundPulmonary Artery Catheter (PAC) knotting is a rare complication of PAC insertion. In patients with dilated right heart chambers, blind insertion of PAC significantly increases the risk of catheter knotting. We demonstrate a safe and successful approach to resolving a PAC knot around pacing leads of a cardiac resynchronization device. Case presentationA 63-year-old African American male with dilated cardiomyopathy and a cardiac resynchronization therapy (CRT) device for severe left ventricular systolic dysfunction required PAC insertion for hemodynamic management of acute heart failure. PAC insertion was complicated by catheter knotting around the pacing leads. The PAC was successfully retrieved using a transvenous technique. ConclusionFluoroscopy-guided insertion of PAC is advisable and preferred over blind insertion in patients with high risk of PAC entanglement. Learning objectiveTo highlight a potential complication of blind pulmonary artery catheter insertion and provide a safe technique to resolve catheter knots.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.