Abstract
Three patients are described in whom pacemaker electrodes were unintentionally placed within the left ventricle, followed by considerable delay before the error was recognized. In two cases temporary pacemaker wires were inserted into the subclavian artery and passed along a retrograde course. One patient required urgent surgery for acute arterial obstruction on removal of the wire. In the third case, a permanent wire was inserted correctly into a vein but traversed the atrial septum, probably via a patent foramen ovale, to enter the left ventricle. Twelve lead electrocardiograms in all three patients showed paced complexes with right bundle branch block configuration. This appearance should raise suspicion that the pacemaker electrode might be in the left ventricle, in which case its position should be defined by chest radiographs (including a lateral view) and echocardiography.
Published Version (Free)
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.