Abstract

Bicaval dual lumen catheters improve the efficiency of veno-venous extracorporeal membrane oxygenation by minimizing recirculation with an innovative design, which requires precise placement of three catheter ports in the superior vena cava, right atrium, and inferior vena cava, respectively. However, the exact position of these catheter ports is usually not known during placement because they cannot be visualized with conventional radiography. We performed a retrospective review of our experience over the past year using transthoracic echocardiography to evaluate the position of the catheter ports. From a subcostal, sagittal imaging approach, we were able to identify all three catheter ports in 11 of 11 studies. At least one of the catheter ports was incorrectly positioned in 5 of 11 studies. Further prospective evaluation is necessary to determine if catheter repositioning based on transthoracic echocardiography findings can further improve the clinical efficiency of veno-venous extracorporeal membrane oxygenation.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.