Abstract

Coronary artery vibrometry is a new transthoracic Doppler ultrasound method for the detection of coronary artery stenosis. It detects audio-frequency vibrations generated by coronary artery luminal diameter reduction. We studied 31 patients with known or suspected stenosis using coronary artery vibrometry and quantitative coronary angiography and 83 normal volunteers. A tissue vibration difference index (TVDI) was calculated from the left anterior descending, circumflex, left main and right coronary arteries. Accuracy for coronary artery stenosis detection using TVDI was assessed. Sensitivity for detecting coronary stenosis equal or greater than 25% diameter reduction was 89% in the left anterior descending coronary artery (16/18, 95% confidence interval [CI] = 64%–98%), 87% in the right coronary artery (13/15, 95% CI = 58%–98%), 83% in the circumflex coronary artery (5/6, 95% CI = 36%–99%) and 100% in the left main artery (3/3, 95% CI = 31%–100%). The median TVDI increased with severity of stenosis, suggesting that this measure might be used to track progression/regression of coronary artery stenosis.(E-mail: kwbeach@u.washington.edu)

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.