Abstract
Existing light sources for intraocular illumination are often bulky and expensive and pose a risk for the patient, because light guides are inserted in the eye through incisions and if the tip of these light guides get too close to the retina, the retina can be damaged photochemically within minutes or even seconds. Therefore a new, safe and simple device for intraocular illumination is developed and evaluated for its thermal and photochemical risks to the patient. It consists of a white LED which is integrated into a pen like holder. This device is pressed against the sclera by the physician who seeks for illumination during surgery or for diagnostic purposes. The LED light is transmitted through the sclera without the need for an incision. Considering the relevant standards, the device poses no harm to the patient, and in tests with the authors' own eyes a sufficient intraocular illumination is reached. The proposed device is quite simple but easy to handle and very gentle for the patient.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.